Pressure transducers DMU 05 P precision version

 SAL 67810 67820 67830 67840 67850 67860 67870 67880 67890 67900 67910 67920 67930 67940 67950 67960 67970 67980 67990 68000 68010 68020 68030 68040 object_image_64527imagemain_en
 SAR 67810 67820 67830 67840 67850 67860 67870 67880 67890 67900 67910 67920 67930 67940 67950 67960 67970 67980 67990 68000 68010 68020 68030 68040
 SAR 67810 67820 67830 67840 67850 67860 67870 67880 67890 67900 67910 67920 67930 67940 67950 67960 67970 67980 67990 68000 68010 68020 68030 68040
 BEF 67810 67820 67830 67840 67850 67860 67870 67880 67890 67900 67910 67920 67930 67940 67950 67960 67970 67980 67990 68000 68010 68020 68030 68040
 BEF 68060 68070 68080 68090 68100 68110 68120 68130 68140 68150 68160 68170 68180 68190
Data sheet Pressure transducers DMU 05 P precision version
AFRISO Data sheet Pressure transducers DMU 05 P precision version

Versions

Type Connection Measuring range Part no. Remember
DMU 05 P G½B EN 837-1/7.3 -1/0 bar 31742
DMU 05 P G½B EN 837-1/7.3 -1/+1.5 bar 31743
DMU 05 P G½B EN 837-1/7.3 -1/+3 bar 31744
DMU 05 P G½B EN 837-1/7.3 -1/+5 bar 31745
DMU 05 P G½B EN 837-1/7.3 0/100 mbar 33026
DMU 05 P G½B EN 837-1/7.3 0/160 mbar 31747
DMU 05 P G½B EN 837-1/7.3 0/250 mbar 31748
DMU 05 P G½B EN 837-1/7.3 0/400 mbar 31749
DMU 05 P G½B EN 837-1/7.3 0/600 mbar 31750
DMU 05 P G½B EN 837-1/7.3 0/1 bar 31751
DMU 05 P G½B EN 837-1/7.3 0/1.6 bar 31752
DMU 05 P G½B EN 837-1/7.3 0/2.5 bar 31753
DMU 05 P G½B EN 837-1/7.3 0/4 bar 31754
DMU 05 P G½B EN 837-1/7.3 0/6 bar 31755
DMU 05 P G½B EN 837-1/7.3 0/10 bar 31756
DMU 05 P G½B EN 837-1/7.3 0/16 bar 31757
DMU 05 P G½B EN 837-1/7.3 0/25 bar 31758
DMU 05 P G½B EN 837-1/7.3 0/40 bar 31759
DMU 05 P G½B EN 837-1/7.3 0/60 bar 31760
DMU 05 P G½B EN 837-1/7.3 0/100 bar 31761
DMU 05 P G½B EN 837-1/7.3 0/160 bar 31762
DMU 05 P G½B EN 837-1/7.3 0/250 bar 31763
DMU 05 P G½B EN 837-1/7.3 0/400 bar 31764
DMU 05 P G½B EN 837-1/7.3 0/600 bar 31765
DMU 05 P VM G½B DIN 3852 type E 0/100 mbar 33027
DMU 05 P VM G½B DIN 3852 type E 0/160 mbar 31771
DMU 05 P VM G½B DIN 3852 type E 0/250 mbar 31772
DMU 05 P VM G½B DIN 3852 type E 0/400 mbar 31773
DMU 05 P VM G½B DIN 3852 type E 0/600 mbar 31774
DMU 05 P VM G½B DIN 3852 type E 0/1 bar 31775
DMU 05 P VM G½B DIN 3852 type E 0/1.6 bar 31776
DMU 05 P VM G½B DIN 3852 type E 0/2.5 bar 31777
DMU 05 P VM G½B DIN 3852 type E 0/4 bar 31778
DMU 05 P VM G½B DIN 3852 type E 0/6 bar 31779
DMU 05 P VM G½B DIN 3852 type E 0/10 bar 31780
DMU 05 P VM G½B DIN 3852 type E 0/16 bar 31781
DMU 05 P VM G½B DIN 3852 type E 0/25 bar 31782
DMU 05 P VM G½B DIN 3852 type E 0/40 bar 33028

Blue part no. = in-stock items

Description Format Language Download / Link
Data sheet Pressure transducers DMU 05 P precision version Data sheet PDF
AFRISO Data sheet Pressure transducers DMU 05 P precision version
Pressure transmitters DMU 03…Ex, DMU 04…Ex, DMU 05…Ex, DMU 07…Ex Operating instructions PDF
AFRISO Pressure transmitters DMU 03…Ex, DMU 04…Ex, DMU 05…Ex, DMU 07…Ex
Pressure transmitters / screw-in probes DMU 01, DMU 01 K, DMU 03, DMU 03 HD, DMU 04, DMU 05P, DMU 07, DMU 07 FG Operating instructions PDF
AFRISO Pressure transmitters / screw-in probes DMU 01, DMU 01 K, DMU 03, DMU 03 HD, DMU 04, DMU 05P, DMU 07, DMU 07 FG
Electronic pressure measurement - pressure transducers DMU series Product range industrial technology PDF
AFRISO Electronic pressure measurement - pressure transducers DMU series
Enquiry pressure transducers DMU Checklists PDF
AFRISO Enquiry pressure transducers DMU
Pressure transducers DMU 01, DMU 03, DMU 04, DMU 05, DMU 08, DMU 08T Declaration of Conformity PDF
AFRISO Pressure transducers DMU 01, DMU 03, DMU 04, DMU 05, DMU 08, DMU 08T
NA Pressure transducers DMU 04 Industrial version Options PDF
AFRISO NA Pressure transducers DMU 04 Industrial version
Pressure transducers convert physical pressure into an electrical signal proportional to the pressure. DMU 05 P is equipped with an oil-filled piezo-resistive silicon measuring cell. The intelligent DMU 05 P pressure transducers are equipped with digital amplifier electronics (microprocessor and 16 bit A/D converter). DMU 05 P actively compensates for sensor-specific deviations (non-linearity and temperature error), allowing for superior measuring characteristics. DMU 05 P can also be supplied with optional digital RS 485 Modbus RTU for setting offset, range and damping.
Measuring accuracy
Deviation from the characteristic curve according to IEC 60770 – limit point calibration (non-linearity, hysteresis, repeatability)
< ± 0.1 % FSO
Measuring range
See ordering table
Overpressure safety
< 600 bar:
At least 2 x FS
≥ 600 bar:
At least 3.5 x FS
Burst pressure
< 400 bar:
At least 5 x FS
≥ 400 bar:
At least 2 x FS
Operating temperature range
Medium:
25/125 °C
Ambient:
-25/+85 °C
Storage:
-40/+100 °C
Temperature error band
≤ ± 0.02 % FSO/10 K In compensated range -20/+80 °C
Dynamic characteristics
Response time:
< 5 ms
Process connection
G½B EN 837-1/7.3
G½B DIN 3852 type E with protruding diaphragm
Material
Housing:
Stainless steel 316 L
Pressure connection:
Stainless steel 316 L
Diaphragm:
Stainless steel 316 L
Seal:
FKM (Viton)
Pressure transmission liquid
Silicone oil
Supply voltage
DC 12 – 36 V
Output signal
2-wire, 0/4–20 mA
Load
R<sub>Max</sub> = [(UB – UB<sub>Min</sub>)/0.02 A] Ω
Current input
< 25 mA
Electrical protection
Short circuit proof and protected against reverse polarity
Electrical connection
Connector and junction box as per ISO 4400 (DIN 43650-A)
Degree of protection
IP 65 (EN 60529)
CE conformity
EMC Directive 2014/30/EU
RoHS Directive 2011/65/EU
Pressure Equipment Directive 2014/68/EU

Options

  • Other process connections
  • Other electrical connections
  • Other seal materials
  • Other output signals
  • RS 232 interface in conjunction with Binder connector 723, 7-pin (interface and software required)
  • EX version (II 1G Ex ia IIC T4 Ga, II 1D Ex ia IIIC T85°C Da)
DMU 05 P - connection G½B EN 837
DMU 05 P - connection G½B EN 837
AFRISO
Dimensions (in mm)
left: version up to 0/40 bar:
right: Version greater than 0/60 bar
DMU 05 P VM - connection G½B DIN 3852 type E
DMU 05 P VM - connection G½B DIN 3852 type E
with protruding diaphragm
AFRISO
Dimensions (in mm)
wiring diagram
wiring diagram
AFRISO
2-wire, 4-20 mA