Display
5-digit graphical LC display,
With backlight,
Text-based user interface,
User interface language selectable (English/German/French/Italian),
Freely selectable units,
Custom display units can be defined
Measuring range
± 99,999 Digits
Start and end values scalable as required
Linearity
± 0.1 % v. Mw.
Resolution
Decimal point position can be set as required
Response time
< 0.2 s
Operating temperature range
Ambient:
0/50 °C
Supply voltage
AC 50 – 253 V, 4.4 VA
DC 20 – 253 V, 2.5 W
Sensor supply
Integrated, galvanically isolated supply voltage for transducer: ≥ DC 21 V at 20 mA
Sensor input
All analogue standard signals, e.g. 4–20 mA, 0–20 mA, 0–1 V, 0–10 V as well as potentiometer
Analogue output
0/4–20 mA, galvanically isolated
Analogue output 2
DA 12:
0–10 V, galvanically isolated
DA 14:
0–10 V, galvanically isolated
Switching output
DA 12:
2 voltage-free changeover contacts
DA 14:
4 voltage-free changeover contacts
Switching hysteresis adjustable
Contact rating:
AC 250 V, 2 A, 100 VA
Housing
Standard rack mounting housing
W x H x D:
96 x 48 x 135 mm
Panel cut out
W x H:
92 x 45 mm
Degree of protection (front)
IP 65 (EN 60529)
Electrical connection
Plug-in screw terminals (1.5 mm<sup>2</sup>)
Linearisation
Customer-specific linearisation with a max. of 24 points for the indication of volume (e.g. litres) in non-linear tanks. Bearing charts for cylindrical horizontal tanks and spherical tanks are pre-programmed.
Min./max. value memory
The highest and lowest values reached during operation can be displayed.